INHECO Single Plate Incubators
INHECO’s single plate incubators heat and protect one plate at a time, with shaking on the shaker variants. Revolution’s driver covers the whole family, and recognises four device types that correspond exactly to INHECO’s own product split:
- Incubator MP — microplate, no shaker
- Incubator DWP — deep-well plate, no shaker
- Incubator Shaker MP — microplate, with shaker
- Incubator Shaker DWP — deep-well plate, with shaker
Which one you have determines whether the shaking figures below apply at all.
At a glance
Section titled “At a glance”| Temperature range | Ambient +5 K to 80 °C |
| Temperature accuracy | 0.3 K |
| Homogeneity | 0.5 K |
| Shaking speed (shaker variants) | 400 – 1,800 rpm |
| Shaking amplitude (shaker variants) | 0 – 3 mm |
| Revolution control | Open and close the door, set temperature, read temperature |
Specifications
Section titled “Specifications”INHECO additionally publishes the following, beyond the figures already in “At a glance” above.
| Maximum temperature | 80 °C |
| Shaking patterns (shaker variants) | Linear, orbital, elliptic and figure-eight, freely programmable |
On heat-up, INHECO states samples go from 20 °C to 37 °C in approximately 20 minutes, and that the Single Plate Incubator Shaker reaches its operating temperature in 10 minutes. Those two figures answer different questions — how long until the samples are at temperature, versus how long until the instrument is — and a schedule that confuses them will start incubating too early.
Note the temperature range is expressed as ambient +5 K at the bottom end. This is a heating device: it cannot go below ambient, so the coldest temperature it can hold depends on the room. If you need to cool below ambient, that is a Thermoshake or a Heater Cooler, not this.
Supported methods
Section titled “Supported methods”- Open door / Close door — move the door so a robot can load and unload
- Set temperature — set the target incubation temperature
- Dump temperature — read the current temperature back
Integration
Section titled “Integration”Revolution talks to the incubator over a serial connection.
Prerequisites
Section titled “Prerequisites”- The unit must be cabled to the Revolution host and its COM port known before the device is initialised.
- The device type must match the hardware — a deep-well unit and a microplate unit take different labware, and the driver distinguishes them.
- The door must be free to open, with the robot’s approach path clear of its travel.
If you need more of this instrument driven from a schedule, get in touch — the driver is extended on demand.